Mooney Check: Test Results
Below you can find the description of the test results produced by MooneyCheck software



 

MU_xx: Mooney Viscosity at set time 
The calculation returns the Mooney Viscosity after the set test time. Standard test time includes 1 minute pre-heating time and 4 minutes test time.

REFERENCE STANDARD: ISO 289-1 / ASTM D 1646

UNIT: MU (Mooney units)

CALCULATION PARAMETERS:
Test Time: t

EXPRESSION:
ML (1+t) - Example ML (1+4)


MU_min: Minimum Mooney Viscosity 
The calculation returns the minimum value of Mooney Viscosity over the complete test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_min




MU_ini: Initial Mooney Viscosity 
The calculation returns the Maximum Mooney viscosity at the beginning of test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_ini


S_xx: Viscosity reduction Rate at set time 
The calculation returns the Slope of curve of the Viscosity vs Time at the set test time.

UNIT: MU (Mooney units)

PARAMETERS:
Test Time: t

EXPRESSION:
S (1+t) - Example S(1+12)


Delta_MU_T1_T2: Variation of Mooney Viscosity  
Reference standard ISO 289-3
The calculation returns the difference between the Mooney viscosities recorded at running time t1 and t2.

UNIT: MU (mooney units)

PARAMETERS:
Test Time t1
Test Time t2

EXPRESSION:
Dt_MU_(t2-t1)




ts: Scorch Time 
OVERVIEW:
The pre-vulcanization characteristics provide a means of estimating how long compounded rubber can be maintained at high temperatures and remain processable.

REFERENCE STANDARD:
ISO 289-3 / ASTM D 1646

The calculation returns the time in minutes including warm-up time, for the viscosity to increase by a specified amount from the minimum value.

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Torque Increase: MU

EXPRESSION:
ts_MU (example ts_35)


Dt_ts: Scorch Index 
The calculation returns the time needed for the viscosity to increase from the first to the second set values.

REFERENCE STANDARD: ASTM D 1646

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Start Torque Value: ts1
Final Torque Value: ts2

EXPRESSION:
Dt_ts_(ts2-ts1) - Example: Dt_ts(35-5)


a (MSR): Mooney Stress-Relaxation Rate 
OVERVIEW:
Stress Relaxation test consists of determining the decay of the Mooney torque immediately after the determination of the Mooney viscosity. After abruptly stopping the rotor at the end of the Mooney viscosity measurement, the decrease in torque is recorded as a function of time. The rate of change of the torque is evaluated over a short time interval.

REFERENCE STANDARD: ISO 289-4

CALCULATION:
The calculation returns the slope of the linear regression line of the log(torque) versus log(time) plot over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement. According to ISO Standard the regression is calculated between 1.6 and 5 seconds from rotor stop.

UNIT: Mu/sec

PARAMETERS:
Time for Start of Regression Calculation: t1
Time for Stop of Regression Calculation: t2

EXPRESSION:
a(MSR) t1_t2 - Example: a(MSR) 1.6_5.0




A: Mooney Stress-Relaxation Area 
The calculation returns the area under the stress relaxation curve over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement.

REFERENCE STANDARD: ASTM D 1646

UNIT: Mu*sec

PARAMETERS:
Time for Start of the Calculation: t1
Time for Stop of the Calculation: t2

EXPRESSION:
A (t1_t2) - Example: A (1.6_5.0)



Gibitre Instruments
Via dell'industria, 18
Bergamo (Italy)
te. +39 035 461146
www.gibitre.it
info@gibitre.it

Copyright Gibitre Instruments
Below you can find the description of the test results produced by MooneyCheck software
Campi di applicazione  >  Prove su pneumatici  >  Rheocheck MD - Drive
Mooney
Check:
Test Results
viscometer mooney formulas: ムーニー粘度計:ソフトウェアの計算式:スコーチタイム。

Below you can find the description of the test results produced by MooneyCheck software

Applications

MU_xx: Mooney Viscosity at set time 

The calculation returns the Mooney Viscosity after the set test time. Standard test time includes 1 minute pre-heating time and 4 minutes test time.

REFERENCE STANDARD: ISO 289-1 / ASTM D 1646

UNIT: MU (Mooney units)

CALCULATION PARAMETERS:
Test Time: t

EXPRESSION:
ML (1+t) - Example ML (1+4)

viscometer mooney formulas: ジビトレソフト。結果計算式Mooney-Checkソフトウェア粘度試験設定時間後のムーニー粘度
ジビトレソフト。結果計算式Mooney-Checkソフトウェア粘度試験設定時間後のムーニー粘度

MU_min: Minimum Mooney Viscosity 

The calculation returns the minimum value of Mooney Viscosity over the complete test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_min

viscometer mooney formulas: ジビトレソフト。結果計算式Mooney-Checkソフトウェア粘度試験。Mu_minの測定:試験中に記録された粘度の最小値。
ジビトレソフト。結果計算式Mooney-Checkソフトウェア粘度試験。Mu_minの測定:試験中に記録された粘度の最小値。

MU_ini: Initial Mooney Viscosity 

The calculation returns the Maximum Mooney viscosity at the beginning of test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_ini

viscometer mooney formulas: ジビトレソフト。結果計算式Mooney-Checkソフトウェア粘度試験。ローターの回転が活性化した瞬間の最大粘度。
ジビトレソフト。結果計算式Mooney-Checkソフトウェア粘度試験。ローターの回転が活性化した瞬間の最大粘度。
 

S_xx: Viscosity reduction Rate at set time 

The calculation returns the Slope of curve of the Viscosity vs Time at the set test time.

UNIT: MU (Mooney units)

PARAMETERS:
Test Time: t

EXPRESSION:
S (1+t) - Example S(1+12)

viscometer mooney formulas: ジビトレソフト。結果計算式Mooney-Checkソフトウェア粘度試験設定時間における粘度曲線の傾きの計算
ジビトレソフト。結果計算式Mooney-Checkソフトウェア粘度試験設定時間における粘度曲線の傾きの計算

Delta_MU_T1_T2: Variation of Mooney Viscosity  

Reference standard ISO 289-3
The calculation returns the difference between the Mooney viscosities recorded at running time t1 and t2.

UNIT: MU (mooney units)

PARAMETERS:
Test Time t1
Test Time t2

EXPRESSION:
Dt_MU_(t2-t1)

viscometer mooney formulas: ジビトレソフト。結果計算式ムーニーチェックソフトウェアデルタムーニー計算:設定時間t1とt2の間の粘度値の変化。
ジビトレソフト。結果計算式ムーニーチェックソフトウェアデルタムーニー計算:設定時間t1とt2の間の粘度値の変化。

ts: Scorch Time 

OVERVIEW:
The pre-vulcanization characteristics provide a means of estimating how long compounded rubber can be maintained at high temperatures and remain processable.

REFERENCE STANDARD:
ISO 289-3 / ASTM D 1646

The calculation returns the time in minutes including warm-up time, for the viscosity to increase by a specified amount from the minimum value.

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Torque Increase: MU

EXPRESSION:
ts_MU (example ts_35)

viscometer mooney formulas: ジビトレソフト。結果計算式ムーニーチェックソフトスコーチテストtsの計算:設定された粘度値に到達するのに必要な時間(例:ts_5)。
ジビトレソフト。結果計算式ムーニーチェックソフトスコーチテストtsの計算:設定された粘度値に到達するのに必要な時間(例:ts_5)。

Dt_ts: Scorch Index 

The calculation returns the time needed for the viscosity to increase from the first to the second set values.

REFERENCE STANDARD: ASTM D 1646

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Start Torque Value: ts1
Final Torque Value: ts2

EXPRESSION:
Dt_ts_(ts2-ts1) - Example: Dt_ts(35-5)

viscometer mooney formulas: ジビトレソフト。結果計算式ムーニーチェックソフトスコーチテストDelta_tsの計算:5~35MUの粘度上昇に要する時間。
ジビトレソフト。結果計算式ムーニーチェックソフトスコーチテストDelta_tsの計算:5~35MUの粘度上昇に要する時間。

a (MSR): Mooney Stress-Relaxation Rate 

OVERVIEW:
Stress Relaxation test consists of determining the decay of the Mooney torque immediately after the determination of the Mooney viscosity. After abruptly stopping the rotor at the end of the Mooney viscosity measurement, the decrease in torque is recorded as a function of time. The rate of change of the torque is evaluated over a short time interval.

REFERENCE STANDARD: ISO 289-4

CALCULATION:
The calculation returns the slope of the linear regression line of the log(torque) versus log(time) plot over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement. According to ISO Standard the regression is calculated between 1.6 and 5 seconds from rotor stop.

UNIT: Mu/sec

PARAMETERS:
Time for Start of Regression Calculation: t1
Time for Stop of Regression Calculation: t2

EXPRESSION:
a(MSR) t1_t2 - Example: a(MSR) 1.6_5.0

viscometer mooney formulas: ジビトレソフト。結果計算式Mooney-Checkソフトウェア。リラクセーションストレス処理の回帰直線を計算するための時間t1とt2の定義。 viscometer mooney formulas: ジビトレソフト。結果計算式Mooney-Checkソフトウェア。リラクセーションストレス処理の回帰直線を計算するための時間t1とt2の定義。 viscometer mooney formulas: Gibitreソフトウェア。結果の計算式。ムーニーチェック。緩和試験の回帰直線の計算式。両対数図での表現。

A: Mooney Stress-Relaxation Area 

The calculation returns the area under the stress relaxation curve over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement.

REFERENCE STANDARD: ASTM D 1646

UNIT: Mu*sec

PARAMETERS:
Time for Start of the Calculation: t1
Time for Stop of the Calculation: t2

EXPRESSION:
A (t1_t2) - Example: A (1.6_5.0)

viscometer mooney formulas: ジビトレソフト。結果計算式Mooney-Checkソフトウェア。応力-弛緩試験におけるエネルギーの計算。
ジビトレソフト。結果計算式Mooney-Checkソフトウェア。応力-弛緩試験におけるエネルギーの計算。
Download
General catalogue mail request

RheoCheck: Test Results
Vedi anche: ソフトウェアレオメーター: ソフトウェア計算式。デルタTnagent
See Also   >
News
Call to action
Mail
info@gibitre.it
The good old e-mail
ready for every season
Phone
+39 035 460146
We welcome your calls Monday through Friday from 8:30 a.m. to 5:30 p.m. in your local time zone.
Newsletter
Subscribe
Stay updated
with the last news
of your sector
Local agent
from: -
See the whole list

In your country
in your language
at your service
Gibitre Instruments S.r.l., Via dell'industria 73, 24126 Bergamo Italia, pec@gibitre-instruments.it P. IVA: IT02827900164 CCIAA REA Bergamo: 326486 | ©2018 - 2026 Gibitre Instruments
Optimist.it
Welcome to the Gibitre Instruments web site
Please chose the preferred Cookie Management.
Benvenuti nel sito Gibitre Instruments
Per continuare scegliete la Gestione dei Cookie che preferite.
Cookies policy
For detailed informations about the use of cookies, please read our
Cookies Management Practice.

Gestione dei Cookies
Per maggiori informazioni sui cookie utilizzati consultate la nostra
Procedura sull'Uso dei Cookies.