Mooney Check: Test Results
Below you can find the description of the test results produced by MooneyCheck software



 

MU_xx: Mooney Viscosity at set time 
The calculation returns the Mooney Viscosity after the set test time. Standard test time includes 1 minute pre-heating time and 4 minutes test time.

REFERENCE STANDARD: ISO 289-1 / ASTM D 1646

UNIT: MU (Mooney units)

CALCULATION PARAMETERS:
Test Time: t

EXPRESSION:
ML (1+t) - Example ML (1+4)


MU_min: Minimum Mooney Viscosity 
The calculation returns the minimum value of Mooney Viscosity over the complete test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_min




MU_ini: Initial Mooney Viscosity 
The calculation returns the Maximum Mooney viscosity at the beginning of test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_ini


S_xx: Viscosity reduction Rate at set time 
The calculation returns the Slope of curve of the Viscosity vs Time at the set test time.

UNIT: MU (Mooney units)

PARAMETERS:
Test Time: t

EXPRESSION:
S (1+t) - Example S(1+12)


Delta_MU_T1_T2: Variation of Mooney Viscosity  
Reference standard ISO 289-3
The calculation returns the difference between the Mooney viscosities recorded at running time t1 and t2.

UNIT: MU (mooney units)

PARAMETERS:
Test Time t1
Test Time t2

EXPRESSION:
Dt_MU_(t2-t1)




ts: Scorch Time 
OVERVIEW:
The pre-vulcanization characteristics provide a means of estimating how long compounded rubber can be maintained at high temperatures and remain processable.

REFERENCE STANDARD:
ISO 289-3 / ASTM D 1646

The calculation returns the time in minutes including warm-up time, for the viscosity to increase by a specified amount from the minimum value.

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Torque Increase: MU

EXPRESSION:
ts_MU (example ts_35)


Dt_ts: Scorch Index 
The calculation returns the time needed for the viscosity to increase from the first to the second set values.

REFERENCE STANDARD: ASTM D 1646

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Start Torque Value: ts1
Final Torque Value: ts2

EXPRESSION:
Dt_ts_(ts2-ts1) - Example: Dt_ts(35-5)


a (MSR): Mooney Stress-Relaxation Rate 
OVERVIEW:
Stress Relaxation test consists of determining the decay of the Mooney torque immediately after the determination of the Mooney viscosity. After abruptly stopping the rotor at the end of the Mooney viscosity measurement, the decrease in torque is recorded as a function of time. The rate of change of the torque is evaluated over a short time interval.

REFERENCE STANDARD: ISO 289-4

CALCULATION:
The calculation returns the slope of the linear regression line of the log(torque) versus log(time) plot over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement. According to ISO Standard the regression is calculated between 1.6 and 5 seconds from rotor stop.

UNIT: Mu/sec

PARAMETERS:
Time for Start of Regression Calculation: t1
Time for Stop of Regression Calculation: t2

EXPRESSION:
a(MSR) t1_t2 - Example: a(MSR) 1.6_5.0




A: Mooney Stress-Relaxation Area 
The calculation returns the area under the stress relaxation curve over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement.

REFERENCE STANDARD: ASTM D 1646

UNIT: Mu*sec

PARAMETERS:
Time for Start of the Calculation: t1
Time for Stop of the Calculation: t2

EXPRESSION:
A (t1_t2) - Example: A (1.6_5.0)



Gibitre Instruments
Via dell'industria, 18
Bergamo (Italy)
te. +39 035 461146
www.gibitre.it
info@gibitre.it

Copyright Gibitre Instruments
Below you can find the description of the test results produced by MooneyCheck software
Campi di applicazione  >  Prove su pneumatici  >  Rheocheck MD - Drive
Mooney
Check:
Test Results
viscometer mooney formulas: 门尼粘度计:软件计算公式:焦烧时间。

Below you can find the description of the test results produced by MooneyCheck software

Applications

MU_xx: Mooney Viscosity at set time 

The calculation returns the Mooney Viscosity after the set test time. Standard test time includes 1 minute pre-heating time and 4 minutes test time.

REFERENCE STANDARD: ISO 289-1 / ASTM D 1646

UNIT: MU (Mooney units)

CALCULATION PARAMETERS:
Test Time: t

EXPRESSION:
ML (1+t) - Example ML (1+4)

viscometer mooney formulas: Gibitre 软件。结果计算公式。Mooney-Check 软件。粘度测试设定时间后的门尼粘度。
Gibitre 软件。结果计算公式。Mooney-Check 软件。粘度测试设定时间后的门尼粘度。

MU_min: Minimum Mooney Viscosity 

The calculation returns the minimum value of Mooney Viscosity over the complete test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_min

viscometer mooney formulas: Gibitre 软件。结果计算公式。Mooney-Check 软件。粘度测试。测量 Mu_min:测试期间记录的最小粘度值。
Gibitre 软件。结果计算公式。Mooney-Check 软件。粘度测试。测量 Mu_min:测试期间记录的最小粘度值。

MU_ini: Initial Mooney Viscosity 

The calculation returns the Maximum Mooney viscosity at the beginning of test time.

UNIT: MU (Mooney units)

EXPRESSION:
MU_ini

viscometer mooney formulas: Gibitre 软件。结果计算公式。Mooney-Check 软件。粘度测试转子旋转启动时的最大粘度。
Gibitre 软件。结果计算公式。Mooney-Check 软件。粘度测试转子旋转启动时的最大粘度。
 

S_xx: Viscosity reduction Rate at set time 

The calculation returns the Slope of curve of the Viscosity vs Time at the set test time.

UNIT: MU (Mooney units)

PARAMETERS:
Test Time: t

EXPRESSION:
S (1+t) - Example S(1+12)

viscometer mooney formulas: Gibitre 软件。结果计算公式。Mooney-Check 软件。粘度测试计算设定时间的粘度曲线斜率。
Gibitre 软件。结果计算公式。Mooney-Check 软件。粘度测试计算设定时间的粘度曲线斜率。

Delta_MU_T1_T2: Variation of Mooney Viscosity  

Reference standard ISO 289-3
The calculation returns the difference between the Mooney viscosities recorded at running time t1 and t2.

UNIT: MU (mooney units)

PARAMETERS:
Test Time t1
Test Time t2

EXPRESSION:
Dt_MU_(t2-t1)

viscometer mooney formulas: Gibitre 软件。结果计算公式。Mooney-Check 软件。Delta-Mooney 计算:设定时间 t1 和 t2 之间粘度值的变化。
Gibitre 软件。结果计算公式。Mooney-Check 软件。Delta-Mooney 计算:设定时间 t1 和 t2 之间粘度值的变化。

ts: Scorch Time 

OVERVIEW:
The pre-vulcanization characteristics provide a means of estimating how long compounded rubber can be maintained at high temperatures and remain processable.

REFERENCE STANDARD:
ISO 289-3 / ASTM D 1646

The calculation returns the time in minutes including warm-up time, for the viscosity to increase by a specified amount from the minimum value.

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Torque Increase: MU

EXPRESSION:
ts_MU (example ts_35)

viscometer mooney formulas: Gibitre 软件。结果计算公式。Mooney-Check 软件灼烧试验计算 ts:达到设定粘度值所需的时间(如 ts_5)。
Gibitre 软件。结果计算公式。Mooney-Check 软件灼烧试验计算 ts:达到设定粘度值所需的时间(如 ts_5)。

Dt_ts: Scorch Index 

The calculation returns the time needed for the viscosity to increase from the first to the second set values.

REFERENCE STANDARD: ASTM D 1646

UNIT: mm:cc (minutes, minutes/100) or mm:ss (minutes, seconds)

PARAMETERS:
Start Torque Value: ts1
Final Torque Value: ts2

EXPRESSION:
Dt_ts_(ts2-ts1) - Example: Dt_ts(35-5)

viscometer mooney formulas: Gibitre 软件。结果计算公式。Mooney-Check 软件灼烧试验计算 Delta_ts:粘度在 5 至 35 MU 之间上升所需的时间。
Gibitre 软件。结果计算公式。Mooney-Check 软件灼烧试验计算 Delta_ts:粘度在 5 至 35 MU 之间上升所需的时间。

a (MSR): Mooney Stress-Relaxation Rate 

OVERVIEW:
Stress Relaxation test consists of determining the decay of the Mooney torque immediately after the determination of the Mooney viscosity. After abruptly stopping the rotor at the end of the Mooney viscosity measurement, the decrease in torque is recorded as a function of time. The rate of change of the torque is evaluated over a short time interval.

REFERENCE STANDARD: ISO 289-4

CALCULATION:
The calculation returns the slope of the linear regression line of the log(torque) versus log(time) plot over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement. According to ISO Standard the regression is calculated between 1.6 and 5 seconds from rotor stop.

UNIT: Mu/sec

PARAMETERS:
Time for Start of Regression Calculation: t1
Time for Stop of Regression Calculation: t2

EXPRESSION:
a(MSR) t1_t2 - Example: a(MSR) 1.6_5.0

viscometer mooney formulas: Gibitre 软件。结果计算公式。Mooney-Check 软件。用于计算松弛应力处理回归线的时间 t1 和 t2 的定义。 viscometer mooney formulas: Gibitre 软件。结果计算公式。Mooney-Check 软件。用于计算松弛应力处理回归线的时间 t1 和 t2 的定义。 viscometer mooney formulas: Gibitre 软件。计算结果的公式。穆尼检查松弛试验回归线计算公式。用双对数图表示。

A: Mooney Stress-Relaxation Area 

The calculation returns the area under the stress relaxation curve over a specified time interval after stopping the rotor at the end of a Mooney viscosity measurement.

REFERENCE STANDARD: ASTM D 1646

UNIT: Mu*sec

PARAMETERS:
Time for Start of the Calculation: t1
Time for Stop of the Calculation: t2

EXPRESSION:
A (t1_t2) - Example: A (1.6_5.0)

viscometer mooney formulas: Gibitre 软件。结果计算公式。Mooney-Check 软件应力-松弛测试中的能量计算。
Gibitre 软件。结果计算公式。Mooney-Check 软件应力-松弛测试中的能量计算。
Download
General catalogue mail request

RheoCheck: Test Results
Vedi anche: 软件流变仪:软件计算公式。Delta Tnagent。
See Also   >
News
Call to action
Mail
info@gibitre.it
The good old e-mail
ready for every season
Phone
+39 035 460146
We welcome your calls Monday through Friday from 8:30 a.m. to 5:30 p.m. in your local time zone.
Newsletter
Subscribe
Stay updated
with the last news
of your sector
Local agent
from: -
See the whole list

In your country
in your language
at your service
Gibitre Instruments S.r.l., Via dell'industria 73, 24126 Bergamo Italia, pec@gibitre-instruments.it P. IVA: IT02827900164 CCIAA REA Bergamo: 326486 | ©2018 - 2026 Gibitre Instruments
Optimist.it
Welcome to the Gibitre Instruments web site
Please chose the preferred Cookie Management.
Benvenuti nel sito Gibitre Instruments
Per continuare scegliete la Gestione dei Cookie che preferite.
Cookies policy
For detailed informations about the use of cookies, please read our
Cookies Management Practice.

Gestione dei Cookies
Per maggiori informazioni sui cookie utilizzati consultate la nostra
Procedura sull'Uso dei Cookies.