Flammability Check
Stainless Steel Cabinet for performing flammability tests of plastic materials, foams and rubber


Standards the instrument complies with:   
ASTM D378;   ASTM D635;   ASTM D3801;   ASTM D4804;   ASTM D4986;   ASTM D5048;   IEC 60695-11-10;   ISO 340;   ISO 1210;   ISO 8030;   ISO 9772;   ISO 9773;   UL 94;   VDE VDE 0471-11-11;    

Gibitre UL 94 flammability testing: plastics, textiles; accurate ASTM D5025 
This stainless steel test booth for UL 94 flame resistance is the ideal tool for consistent evaluation of the combustion behavior of materials.

An ASTM D5025-compliant Bunsen burner with 0°, 20°, and 45° adjustable stand, horizontal slide, and pneumatic flame-to-sample distance delivers precise, repeatable positioning. The flow meter provides accurate gas flow measurement, while the digital thermometer with copper probe ensures reliable flame temperature measurement. A digital timer (0.1 s resolution) enables exact control of exposure times.

Applications: UL 94 testing of horizontal HB and HBF specimens and vertical VB and MVB specimens, for solid products and cellular polymeric materials. A tempered glass viewing window and LED lighting aid observation; the exhaust fan removes combustion fumes to maintain clear visibility.

Why choose it: assured standards compliance, full control of critical parameters, and high repeatability with integrated tools and measuring instruments, enablin

Key Features 
Instrument for flame resistance test according to UL 94 standard fitted with:
• Stainless steel cabinet with tempered glass viewing window and interior light
• Bunsen burner according to ASTM D5025 standard
• Flow meter for gas flow regulation
• Support for flame angle set (0°, 20°, 45°)
• Slide for the regulation of the position of the flame
• Sample holder according to the standards with sample position regulation
• Digital thermometer for the measurement of the temperature of the flame integrated in the instrument
• Digital Timer with 0.1 second resolution for the measurement of test time
• Exhaust fan to remove combustion fumes


Displacement of the Burner 
The ability to move the burner in a controlled manner allows it to be positioned exactly under the specimen and to follow its movement during the flame exposure time.
The instrument is equipped with a pneumatically operated slide that is activated by means of a lever located in the front the instrument.
The device allows the operator to perfectly control the position of the flame under the specimen through the inspection window and to obtain highly repeatable test results.




Vertical Displacement of the sample 
UL 94 requires that the distance between the flame and the specimen be kept constant during exposure of the specimen to the flame.
The distance between the flame and the specimen changes during exposure to the flame due to combustion and the tendency of the specimen to bend.
To enable the operator to correct the distance between the flame and the sample, the instrument has been equipped with a pneumatic slide that allows the sample to be moved vertically. The pneumatic slide is operated by a lever located in the front of the instrument.
This unique feature ensures an important improvement in the repeatability of results and compliance with regulatory requirements.


Control Panel 
The control panel includes:
• Digital reading of the temperature of the flame
• Timer for the measure of test time
• Button for Start / Stop button for the Timer
• Button for Reset of the Timer
• Button for activation of internal light
• Button for activation of fume aspiration


Gas Flow Adjustment 
The instrument must be connected to a Methan source.

The instrument is equipped with:
• Manometer between between 0 and 25 mBar
• Fluxometer between 0,04-0,36 NL\min with flow regulation knob
• Fluxometer between 0,2-1,3 NL\min with flow regulation knob




Bunsen Burner  
Bunsen burner compliant with ASTM D5025 and CEI IEC 60695-11-4 standards equipped with:
• Safety sensor for interruption of gas flow with flame absent
• Ring for the adjustment of the air / gas mix
• Tiltable support for testing with inclined flame (0 °, 20 °, 45 °)


Temperature Calibration  
The instrument is supplied with:
• Internal Thermometer with thermocouple for 20 mm Test Flame according to ASTM D 5207 standard.
• Ruler for the check of the height of the flame.

The Thermocuple and the ruler are installed inside the chamber.
To test the temperature of the flame, you only need to move the burner under the Thermocouple using the movement knob.


Air Flow and Fume Extraction  
The instrument is equipped with a manual valve for interrupting the air flow during the test and an aspirator for removing the fumes at the end of the test.



Positioning and Inspection of the Specimen 
The instrument is equipped with sliding guides in the front and side positions that allow the specimen holders to be placed in the optimum position and their behavior during the test to be inspected through the front window.
The mirror, equipped with a telescopic arm, allows inspection of the specimen behavior at the rear as well.
The interior lighting, via LED lights, is activated by a button on the control panel.


Test with Vertical Sample  
Specimen support for Vertical Burning (VB) test conforming to JAR 25 853 B standard.

Testing of Vertical Rigid Sample according to UL 94 (VB) and ASTM D3801  
Specimen support for Vertical Burning of Solid Plastics (VB) test conforming to UL 94 (tests V-0, V-1, V-2, 5VA, 5VB), ASTM D3801 and equivalent standards.



Testing with non-rigid vertical test specimen according to UL 94 (MVB) 
Specimen support for Vertical Burning of Non-rigid Solid Plastics (MVB) test conforming to UL 94 (tests VTM-0, VTM-1 and VTM- 2), ASTM D4804 and equivalent standards.


Test with Horizontal sample 
• Specimen support for Horizontal Burning (HB) test conforming to UL 94, ASTM D 635 and equivalent standards.

Test with Horizontal Sample of Expanded products 
Specimen support for Horizontal Burning of Cellular Polymeric materials (HBF) test conforming to UL 94, ASTM D 4986 and equivalent.



Flame resistance tests of conveyor belts in accordance with ISO 340 and ASTM D378. 
Instrument configuration for performing flame resistance tests of conveyor belts in accordance with ISO 340 and ASTM D378.

The configuration includes:
- Specimen holder with dimensions to standard
- Blowing device with adjustable air flow
- Button for switching the blowing device on and off


 

Instrument Structure  
The instrument, completeley made of Stainless Steel, is optimized for being placed on the table of your lab and for the connection to fume evacuation systems.



Development and production 
The instrument is totally developed and produced in the plant of Gibitre Instruments in Italy.
All the mechanical parts are produced in the company workshop using modern CNC machines.
Components and sensors from well-known brands are selected in order to ensure the maximum reliability in the measures
Internal trained personnel takes care of all the production stages: assembly, start-up, calibration, packing, shipment and installation.



Testing Configuration 
 
Types of test that can be performed 
HBF (Horizontal Burning Foamed Material), HB (Horizontal Burning), MVB (Material Vertical Burning), VB (Vertical Burning) 
Test Cabinet 
Stainless steel with a volume of 0.5 m3 with inspection window and light for internal illumination. 
Burner 
Conforming to ASTM D5025 and CEI IEC 60695-11-4 standards. With support for flame angle set (0°, 20°, 45°). 
Temperature Control 
Internal Digital thermometer in conformity with ASTM D 5207 standard 
Gas flow Regulation (Methane) 
Integrated Flow meter 
Timer 
Digital Timer with 0.1 sec. resolution. The timer permits to set the flame exposition time. At the end of the count down for flame exposition, the timer permits to count flame extinction time. 
Slide for the displacement of the Burner during the test 
The slide, controlled by a pneumatic cylinder, allows the burner to be moved from outside the chamber via the control knob. 
Slide for the regulation of the vertical position of the sample 
The slide, which is controlled by a pneumatic cylinder, allows the distance between the specimen and the flame to be controlled during the execution of the test using the adjustment knob. 
Aspiration of Fumes 
Exhaust fan to remove fumes at the end of the test 
Safety Devices 
 
Safety Devices 
Safety valve for gas flow interruption without flame 
Labelling 
CE Labelling 
Calibration 
 


Calibration 
Calibration Report conforming to UL 94 with traceability to primary standards 
Calibration of test Flame 
Internal Thermometer with thermocouples for 20 mm Test Flame according to ASTM D 5207 standard. 
Technical specifications 
 
Power 
220 VAC ±10%,50 Hz ±3,4 A,single phase, 40 W - 110 VAC ±10%, 60 Hz ±3 on request 
Gas Supply 
The instrument must be connected to a Methane Cylinder.
A pressure reducer (not included) is required to supply the instrument with a pressure of 1 bar.
The connection to the gas network must be carried out by a qualified technician. 
Internal Dimension 
(W x D x H) 850 x 700 x 812 mm 
External Dimension 
(W x D x H) 925 x 800 x 1235 mm 
Weight 
92 kg 


Gibitre Instruments
Via dell'industria, 18
Bergamo (Italy)
te. +39 035 461146
www.gibitre.it
info@gibitre.it

Copyright Gibitre Instruments
Stainless Steel Cabinet for performing flammability tests of plastic materials, foams and rubber
Campi di applicazione  >  Prove su pneumatici  >  Rheocheck MD - Drive
Flammability
Check
flammability ul94: UL94に準拠した難燃性試験キャビン。計器の全景。
Get a Quotation

Stainless Steel Cabinet for performing flammability tests of plastic materials, foams and rubber

Standards the instrument complies with:

Cellular Products
Cellular plastics -- Determination of horizontal burning characteristics of small specimens subjected to a small flameISO 9772Standard Test Method for Horizontal Burning Characteristics of Cellular Polymeric MaterialsASTM D4986
Conveyor Belts
Conveyor belts Laboratory scale flammability characteristics Requirements and test methodISO 340Standard Test Methods for Rubber (Elastomeric) Conveyor Belting, Flat TypeASTM D378
Plastics
Standard for Tests for Flammability of Plastic Materials for Parts in Devices and AppliancesUL 94Plastics — Determination of the burning behaviour of horizontal and vertical specimens in contact with a small-flame ignition sourceISO 1210Plastics -- Determination of burning behaviour of thin flexible vertical specimens in contact with a small-flame ignition sourceISO 9773Standard Test Method for Rate of Burning and/or Extent and Time of Burning of Plastics in a Horizontal PositionASTM D635Standard Test Method for Measuring the Comparative Burning Characteristics of Solid Plastics in a Vertical PositionASTM D3801Standard Test Method for Determining the Flammability Characteristics of Nonrigid Solid PlasticsASTM D4804Standard Test Method for Measuring the Comparative Burning Characteristics and Resistance to Burn-Through of Solid Plastics Using a 125-mm FlameASTM D5048Fire hazard testing — Part 11-10: Test flames — 50 W horizontal and vertical flame test methodsIEC 60695-11-10Fire hazard testing Part 11-11: Test flames – Determination of the characteristic heat flux for ignition from a non-contacting flame source (IEC 60695-11-11:2021)VDE VDE 0471-11-11
Rubber Hoses
Rubber and plastics hoses — Method of test for flammabilityISO 8030

Applications

This stainless steel test booth for UL 94 flame resistance is the ideal tool for consistent evaluation of the combustion behavior of materials.

An ASTM D5025-compliant Bunsen burner with 0°, 20°, and 45° adjustable stand, horizontal slide, and pneumatic flame-to-sample distance delivers precise, repeatable positioning. The flow meter provides accurate gas flow measurement, while the digital thermometer with copper probe ensures reliable flame temperature measurement. A digital timer (0.1 s resolution) enables exact control of exposure times.

Applications: UL 94 testing of horizontal HB and HBF specimens and vertical VB and MVB specimens, for solid products and cellular polymeric materials. A tempered glass viewing window and LED lighting aid observation; the exhaust fan removes combustion fumes to maintain clear visibility.

Why choose it: assured standards compliance, full control of critical parameters, and high repeatability with integrated tools and measuring instruments, enablin
Why Choose Flammability Check ?

Key Features 

Instrument for flame resistance test according to UL 94 standard fitted with:
Stainless steel cabinet with tempered glass viewing window and interior light
Bunsen burner according to ASTM D5025 standard
Flow meter for gas flow regulation
• Support for flame angle set (0°, 20°, 45°)
Slide for the regulation of the position of the flame
• Sample holder according to the standards with sample position regulation
Digital thermometer for the measurement of the temperature of the flame integrated in the instrument
Digital Timer with 0.1 second resolution for the measurement of test time
Exhaust fan to remove combustion fumes

flammability ul94: UL94に準拠した難燃性試験キャビン。計器の全景。
UL94に準拠した難燃性試験キャビン。計器の全景。

Displacement of the Burner 

The ability to move the burner in a controlled manner allows it to be positioned exactly under the specimen and to follow its movement during the flame exposure time.
The instrument is equipped with a pneumatically operated slide that is activated by means of a lever located in the front the instrument.
The device allows the operator to perfectly control the position of the flame under the specimen through the inspection window and to obtain highly repeatable test results.

flammability ul94: UL 94に準拠した火炎試験装置。バーナーの位置決めに空気圧スライドを使用する場合の模式図。 flammability ul94: UL 94に準拠した火炎試験装置。バーナーの位置決めに空気圧スライドを使用する場合の模式図。 flammability ul94: UL 94に準拠した火炎試験装置。キャビン内のバーナーの水平移動図。 flammability ul94: 難燃性-UL94の難燃性測定用。空気圧で試料を上下させるノブ。

Vertical Displacement of the sample 

UL 94 requires that the distance between the flame and the specimen be kept constant during exposure of the specimen to the flame.
The distance between the flame and the specimen changes during exposure to the flame due to combustion and the tendency of the specimen to bend.
To enable the operator to correct the distance between the flame and the sample, the instrument has been equipped with a pneumatic slide that allows the sample to be moved vertically. The pneumatic slide is operated by a lever located in the front of the instrument.
This unique feature ensures an important improvement in the repeatability of results and compliance with regulatory requirements.

flammability ul94: UL 94に準拠した火炎試験装置。試験片の垂直位置決め用空気圧スライドの使用概略図。 flammability ul94: UL 94に準拠した火炎試験装置。試験片の垂直位置決め用空気圧スライドの使用概略図。 flammability ul94: 難燃性-UL94の難燃性測定用。空気圧でバーナーを動かすノブ。
 
Testing Control

Control Panel 

The control panel includes:
• Digital reading of the temperature of the flame
• Timer for the measure of test time
• Button for Start / Stop button for the Timer
• Button for Reset of the Timer
• Button for activation of internal light
• Button for activation of fume aspiration

flammability ul94: UL 94に準拠した難燃性試験キャビン。計器の全景。 flammability ul94: UL 94に準拠した難燃性試験キャビン。計器の全景。 flammability ul94: UL94に準拠した難燃性試験装置。試験ブースの右側にあるコントロールパネル。 flammability ul94: UL94に準拠した難燃性試験装置。試験ブースの右側にある操作パネルの詳細。

Gas Flow Adjustment 

The instrument must be connected to a Methan source.

The instrument is equipped with:
• Manometer between between 0 and 25 mBar
• Fluxometer between 0,04-0,36 NL\min with flow regulation knob
• Fluxometer between 0,2-1,3 NL\min with flow regulation knob

flammability ul94: UL 94に準拠した難燃性試験装置。ガス流量を正確に調整するための流量計。 flammability ul94: UL 94に準拠した難燃性試験装置。ガス流量を正確に調整するための流量計。 flammability ul94: UL94に準拠した難燃性試験装置。ガス流量調整器の詳細 flammability ul94: UL 94火炎試験装置ガス供給用メタンガスボンベ

Bunsen Burner  

Bunsen burner compliant with ASTM D5025 and CEI IEC 60695-11-4 standards equipped with:
• Safety sensor for interruption of gas flow with flame absent
• Ring for the adjustment of the air / gas mix
Tiltable support for testing with inclined flame (0 °, 20 °, 45 °)

flammability ul94: 耐炎試験キャビネット用ブンゼンバーナー詳細 flammability ul94: 耐炎試験キャビネット用ブンゼンバーナー詳細 flammability ul94: 耐炎試験ブース用バーナー詳細 flammability ul94: 燃焼性試験機。ASTM D5025に準拠したバーナー。ブンゼンバーナーの詳細 flammability ul94: 燃焼性-UL94:垂直試験片(VB)を用いた耐炎性試験のためのブンゼンバーナー位置詳細。

Temperature Calibration  

The instrument is supplied with:
• Internal Thermometer with thermocouple for 20 mm Test Flame according to ASTM D 5207 standard.
• Ruler for the check of the height of the flame.

The Thermocuple and the ruler are installed inside the chamber.
To test the temperature of the flame, you only need to move the burner under the Thermocouple using the movement knob.

flammability ul94: UL 94規格に基づく難燃性試験用 Flammability-Check 装置。本装置には規格準拠の温度センサーが内蔵されており、炎の適切な温度を即時確認できます。 flammability ul94: UL 94規格に基づく難燃性試験用 Flammability-Check 装置。本装置には規格準拠の温度センサーが内蔵されており、炎の適切な温度を即時確認できます。 flammability ul94: UL94規格に準拠した耐炎試験器。炎の温度を測定する温度計と、炎の露出時間と消火時間をカウントするタイマーを備えたコントロールパネルの詳細。

Air Flow and Fume Extraction  

The instrument is equipped with a manual valve for interrupting the air flow during the test and an aspirator for removing the fumes at the end of the test.

flammability ul94: 可燃性試験装置には手動開閉バルブが装備されており、試験中にチャンバー内の空気の流れを遮断(UL 94規格の要件)し、試験終了時にバルブを開いて煙を排出することが可能です。 flammability ul94: 可燃性試験装置には手動開閉バルブが装備されており、試験中にチャンバー内の空気の流れを遮断(UL 94規格の要件)し、試験終了時にバルブを開いて煙を排出することが可能です。 flammability ul94: 可燃性-UL 94に従った難燃性試験をチェックする。吸引システムへの空気流を遮断するためのバルブ。 flammability ul94: UL 94に準拠した難燃性試験キャビン。計器の全景。

Positioning and Inspection of the Specimen 

The instrument is equipped with sliding guides in the front and side positions that allow the specimen holders to be placed in the optimum position and their behavior during the test to be inspected through the front window.
The mirror, equipped with a telescopic arm, allows inspection of the specimen behavior at the rear as well.
The interior lighting, via LED lights, is activated by a button on the control panel.

flammability ul94: UL 94に準拠した難燃性試験装置。試験中に試験片の裏側を見るために試験ブースに取り付けられた調整可能な鏡の詳細。 flammability ul94: UL 94に準拠した難燃性試験装置。試験中に試験片の裏側を見るために試験ブースに取り付けられた調整可能な鏡の詳細。 flammability ul94: 難燃性試験装置:試験片の後方に設置できるミラーにより、消炎性を確認できる。
Available Sample Holders

Test with Vertical Sample  

Specimen support for Vertical Burning (VB) test conforming to JAR 25 853 B standard.

flammability ul94: UL 94に準拠した火炎試験装置JAR 25 853 Bに準拠した垂直試験片付き試験スタンド
UL 94に準拠した火炎試験装置JAR 25 853 Bに準拠した垂直試験片付き試験スタンド

Testing of Vertical Rigid Sample according to UL 94 (VB) and ASTM D3801  

Specimen support for Vertical Burning of Solid Plastics (VB) test conforming to UL 94 (tests V-0, V-1, V-2, 5VA, 5VB), ASTM D3801 and equivalent standards.

flammability ul94: UL 94に準拠した難燃性試験用キャビン。UL 94 (VB)およびASTM D3801に準拠した垂直剛性試験スタンド。 flammability ul94: UL 94に準拠した難燃性試験用キャビン。UL 94 (VB)およびASTM D3801に準拠した垂直剛性試験スタンド。 flammability ul94: UL 94に準拠した難燃性試験装置。硬質試験片を用いたVB(垂直燃焼)試験の図。

Testing with non-rigid vertical test specimen according to UL 94 (MVB) 

Specimen support for Vertical Burning of Non-rigid Solid Plastics (MVB) test conforming to UL 94 (tests VTM-0, VTM-1 and VTM- 2), ASTM D4804 and equivalent standards.

flammability ul94: UL 94に準拠した火炎試験ブースUL 94(MVB)に準拠した非剛性垂直試験片を備えた試験スタンド。 flammability ul94: UL 94に準拠した火炎試験ブースUL 94(MVB)に準拠した非剛性垂直試験片を備えた試験スタンド。 flammability ul94: UL 94に準拠した難燃性試験装置。MVB(垂直燃焼)試験スキーム。 flammability ul94: UL 94に準拠した難燃性試験装置。MVB(垂直燃焼)試験の詳細。

Test with Horizontal sample 

• Specimen support for Horizontal Burning (HB) test conforming to UL 94, ASTM D 635 and equivalent standards.

flammability ul94: UL 94に準拠した火炎試験装置UL 94、ASTM D 4986に準拠したエキスパンド素材の水平試験片(HBF)による難燃性試験用メッシュ付き試験片ホルダー。 flammability ul94: UL 94に準拠した火炎試験装置UL 94、ASTM D 4986に準拠したエキスパンド素材の水平試験片(HBF)による難燃性試験用メッシュ付き試験片ホルダー。 flammability ul94: 燃焼性-UL94に従った難燃性試験のチェック。硬質試験片を用いたHB(水平燃焼)試験の模式図

Test with Horizontal Sample of Expanded products 

Specimen support for Horizontal Burning of Cellular Polymeric materials (HBF) test conforming to UL 94, ASTM D 4986 and equivalent.

flammability ul94: UL 94に準拠した難燃性試験装置。UL 94およびASTM D4986に準拠した発泡材料の水平試験片(HBF)による難燃性試験用メッシュ付き試験片ホルダー。 flammability ul94: UL 94に準拠した難燃性試験装置。UL 94およびASTM D4986に準拠した発泡材料の水平試験片(HBF)による難燃性試験用メッシュ付き試験片ホルダー。 flammability ul94: 燃焼性試験機:UL94による発泡製品の水平試験。 flammability ul94: UL 94に準拠した難燃性試験装置。HBF(泡の水平燃焼)試験の模式図。 flammability ul94: UL 94に準拠した難燃性試験装置。HBF(泡の水平燃焼)試験の詳細。

Flame resistance tests of conveyor belts in accordance with ISO 340 and ASTM D378. 

Instrument configuration for performing flame resistance tests of conveyor belts in accordance with ISO 340 and ASTM D378.

The configuration includes:
- Specimen holder with dimensions to standard
- Blowing device with adjustable air flow
- Button for switching the blowing device on and off

flammability ul94: UL-94に準拠した難燃性試験用器具。ISO 340規格に準拠した試験用サポート。 flammability ul94: UL-94に準拠した難燃性試験用器具。ISO 340規格に準拠した試験用サポート。 flammability ul94: UL94に準拠した難燃性試験装置。ISO 340に準拠した試験スタンドの詳細。 flammability ul94: UL94に準拠した難燃性試験装置。ISO 340に準拠した試験スタンドの詳細。 flammability ul94: UL94に準拠した難燃性試験装置。ISO 340に従った試験で使用される空気流を作動させる制御装置の詳細。

 

Instrument Structure  

The instrument, completeley made of Stainless Steel, is optimized for being placed on the table of your lab and for the connection to fume evacuation systems.

flammability ul94: UL 94に準拠した難燃性試験チャンバー。チャンバーの寸法を示す正面図。 flammability ul94: UL 94に準拠した難燃性試験チャンバー。チャンバーの寸法を示す正面図。 flammability ul94: 可燃性チェック:キャビンの寸法。側面図。

Development and production 

The instrument is totally developed and produced in the plant of Gibitre Instruments in Italy.
All the mechanical parts are produced in the company workshop using modern CNC machines.
Components and sensors from well-known brands are selected in order to ensure the maximum reliability in the measures
Internal trained personnel takes care of all the production stages: assembly, start-up, calibration, packing, shipment and installation.

flammability ul94: 誇り高きイタリア製
誇り高きイタリア製

Technical specification

Reference Standards
Standards the instrument complies with:  ISO 9772; ASTM D4986; ISO 340; ASTM D378; UL 94; ISO 1210; ISO 9773; ASTM D635; ASTM D3801; ASTM D4804; ASTM D5048; IEC 60695-11-10; VDE VDE 0471-11-11; ISO 8030;  

Testing Configuration

Types of test that can be performed  HBF (Horizontal Burning Foamed Material), HB (Horizontal Burning), MVB (Material Vertical Burning), VB (Vertical Burning)  Test Cabinet  Stainless steel with a volume of 0.5 m3 with inspection window and light for internal illumination.  Burner  Conforming to ASTM D5025 and CEI IEC 60695-11-4 standards. With support for flame angle set (0°, 20°, 45°).  Temperature Control  Internal Digital thermometer in conformity with ASTM D 5207 standard  Gas flow Regulation (Methane)  Integrated Flow meter  Timer  Digital Timer with 0.1 sec. resolution. The timer permits to set the flame exposition time. At the end of the count down for flame exposition, the timer permits to count flame extinction time.  Slide for the displacement of the Burner during the test  The slide, controlled by a pneumatic cylinder, allows the burner to be moved from outside the chamber via the control knob.  Slide for the regulation of the vertical position of the sample  The slide, which is controlled by a pneumatic cylinder, allows the distance between the specimen and the flame to be controlled during the execution of the test using the adjustment knob.  Aspiration of Fumes  Exhaust fan to remove fumes at the end of the test 

Safety Devices

Safety Devices  Safety valve for gas flow interruption without flame  Labelling  CE Labelling 

Calibration

Calibration  Calibration Report conforming to UL 94 with traceability to primary standards  Calibration of test Flame  Internal Thermometer with thermocouples for 20 mm Test Flame according to ASTM D 5207 standard. 

Technical specifications

Power  220 VAC ±10%,50 Hz ±3,4 A,single phase, 40 W - 110 VAC ±10%, 60 Hz ±3 on request  Gas Supply  The instrument must be connected to a Methane Cylinder.
A pressure reducer (not included) is required to supply the instrument with a pressure of 1 bar.
The connection to the gas network must be carried out by a qualified technician. 
Internal Dimension  (W x D x H) 850 x 700 x 812 mm  External Dimension  (W x D x H) 925 x 800 x 1235 mm  Weight  92 kg 

FAQ

References
Gibitre Flammability Check
is currently installed in
18 countries
Request the List of top References
Download
Brief product leaflet
General catalogue mail request

Calibration certificate
example
CE certificate
example
Block Oven Aging Check
Vedi anche: ISO 1817およびASTM D471に準拠した促進老化試験用温度制御ブロック
See Also   >
Multi-Head Die Cutter
Vedi anche: 試料を準備するための金型を適用する手動型抜き機。
See Also   >
News
Call to action
Ask for quotation
a guided way to ask
for a quotation
Mail
info@gibitre.it
The good old e-mail
ready for every season
Phone
+39 035 460146
We welcome your calls Monday through Friday from 8:30 a.m. to 5:30 p.m. in your local time zone.
Newsletter
Subscribe
Stay updated
with the last news
of your sector
Local agent
from: -
See the whole list

In your country
in your language
at your service
Gibitre Instruments S.r.l., Via dell'industria 73, 24126 Bergamo Italia, pec@gibitre-instruments.it P. IVA: IT02827900164 CCIAA REA Bergamo: 326486 | ©2018 - 2026 Gibitre Instruments
Optimist.it
Welcome to the Gibitre Instruments web site
Please chose the preferred Cookie Management.
Benvenuti nel sito Gibitre Instruments
Per continuare scegliete la Gestione dei Cookie che preferite.
Cookies policy
For detailed informations about the use of cookies, please read our
Cookies Management Practice.

Gestione dei Cookies
Per maggiori informazioni sui cookie utilizzati consultate la nostra
Procedura sull'Uso dei Cookies.